| Model Number |
1457Q/75 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that the patient has deceased on an unknown date.There is no known allegation from a health care professional that suggests that the death was device related.The cause of death was unknown.No additional information was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The lead was returned due to a patient death.As received, only the connector portion of the lead was returned in one piece for analysis.Visual examination of the partial lead did not find any anomalies except for procedural damage.Electrical testing did not find any indication of conductor fractures or internal shorts.
|
| |
|