• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL (CRM-SYLMAR) ENTRANT VR; No Match

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL (CRM-SYLMAR) ENTRANT VR; No Match Back to Search Results
Model Number CDVRA300Q
Medical Device Problem Codes Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
Following a magnetic resonance imaging (mri), the device as found to be in backup vvi mode (bvvi).The device has not been programmed into mri mode.Technical support was contacted and the device was reprogrammed to resolve the event.The patient was stable with no consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRANT VR
Common Device Name
No Match
Manufacturer (Section D)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT MEDICAL (CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key25037787
Report Number2017865-2026-08824
Device Sequence Number13376539
Product Code NIK
UDI-Device Identifier05415067032836
UDI-Public(01)05415067032836(10)P000135526(17)231130
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model NumberCDVRA300Q
Device Catalogue NumberCDVRA300Q
Device Lot NumberP000135526
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2026
Initial Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-