A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A trilogy 100 ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.The device failed functional testing relating to the positive flow verification test step.The software was upgraded to address the issue.Additionally, the device failed a not-reportable test step relating to the nurse call.The device was checked again, and the nurse call was functional.Also, a not-reportable internal battery depleted, and low expiratory pressure error code were found in the device's error log.The service technician states that the internal battery was charged to resolve the issue.Run-in testing was performed again and the device passed.
|