| Model Number |
IT1054096B |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|
| |
|
Event or Problem Description
|
|
A device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.A trilogy 100 ventilator was returned to a third-party service center for foam remediation on the device.There was no harm or injury reported.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation.The device failed functional testing relating to the positive flow verification test step.The software was upgraded to address the issue.Run-in testing was performed again and the device passed.Additionally, a not-reportable internal battery depleted and check circuit error code were found in the device's error log.The service technician states that the internal battery was charged and the tubing was reconnected to resolve the issues.
|
| |
|
Search Alerts/Recalls
|