(b)(4).D10- medical product persona femur cemented plus right size 7+ item# 42504606212 lot# 64767893.Persona stem ext straight spld uncemented 16mm dia +135mm l item# 42560113516 lot# 64783575.Persona femoral distal augment cemented size 7, 7+ 5mm item# 42556606205 lot# 64859431.Persona tibia fixed cemented right size d item# 42542006702 lot# 64565801.Persona stem ext 3mm offset spld uncemented 12mm dia +135mm l item# 42560313512 lot# 64675872.Persona articular surface fixed bearing (cps) right 16mm item# 42522600516 lot# 64647943.Persona femoral central cone size small item# 42545001011 lot#64643091.Nexgen complete knee solution trabecular metal std patella item# 00587806532 lot# 64227939.Unk palacos cement.H10: this device was erroneously reported under an incorrect mfr, and is now being submitted under the appropriate mfr.See original report 0001822565-2024-01322.No products were returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Office notes were provided and report pain, swelling and trouble ambulating.The patient also reported fall on ice.X-rays taken during visit state no periprosthetic fracture and implants are well positioned.X-rays reviewed over two months during which shows a stable appearance of revision components.The patient reports increasing right knee pain resulting in physical therapy.Review of the device's history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.This complaint is not confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|