• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CORAIL; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CORAIL; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number 3L92503
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Date of Event 04/02/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
 
Event or Problem Description
It was reported the patient had a history of a metal-metal articulating total hip arthroplasty of the right hip.On (b)(6) 2024, an mri of the right hip revealed the presence of fluid collection emanating from the joint.Laboratory evaluation demonstrated elevated plasma metal ion levels.On (b)(6) 2026, the patient underwent revision surgery of the right total hip arthroplasty due to mechanical failure secondary to metal hypersensitivity, resulting in an adverse local tissue reaction (altr).Doi: (b)(6) 2009.Dor: (b)(6) 2026.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint: (b)(4).Investigation summary: no device associated with this report was received for examination.According to the information received, "the patient has a history of a metal-metal articulating total hip arthroplasty of the right hip.On (b)(6) 2024, patient performed an mri of the right hip.The scan reveals the presence of fluid collection measuring 6.5 cm emanating from the joint.Laboratory evaluation demonstrated elevated plasma metal ion levels, with chromium at 9 and cobalt at 10.On (b)(6) 2026, the patient underwent revision surgery of the right total hip arthroplasty due to mechanical failure secondary to metal hypersensitivity, resulting in an adverse local tissue reaction (altr).Doi: (b)(6) 2009.Dor: (b)(6) 2026.Affected side: right hip." product description: corail amt collar size 13.Product code: 3l92503.Lot no: 2774526.Quantity of manufactured: (b)(4).Date of manufacturing: 12-dec-2008.Expiry date: 01-nov-2013.A manufacturing record evaluation was performed for the finished device product description: corail amt collar size 13, product code: 3l92503, lot number: 2774526 and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 12-dec-2008.3) any anomalies or deviations identified in dhr: none.4) expiry date: 01-nov-2013.Device history review: a manufacturing record evaluation was performed for the finished device product description: corail amt collar size 13, product code: 3l92503, lot number: 2774526 and no non-conformances / manufacturing irregularities were identified.Corrected: d4 (expiration date, primary udi #), g4 pma, h4.Added: d10 concomitant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key25046418
Report Number1818910-2026-07972
Device Sequence Number15932884
Product Code LZO
UDI-Device Identifier10603295168690
UDI-Public(01)10603295168690(17)131101(10)2774526
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K042992
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/01/2013
Device Catalogue Number3L92503
Device Lot Number2774526
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/16/2026
Supplement Date Received by Manufacturer05/28/2026
Initial Report FDA Received Date04/30/2026
Supplement Report FDA Received Date05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/12/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12/14 ARTICUL 40MM M SPEC+8.5; APEX HOLE ELIM POSITIVE STOP; PINNACLE 300 ACET CUP 56MM; PINNACLE MTL INS NEUT40IDX56OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
Patient RaceWhite
-
-