The alleged event is a risk associated with breast surgery and there is no evidence to suggest a link between this particular implant and/or its manufacturing process and a higher risk of occurrence.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: breast implant extrusion, or breast implant exposure, occurs when the breast skin and tissues that are holding the implant fail, causing the implant to protrude through the skin and become exposed.It happens in less than 2% of patients; it is shortly after breast augmentation or down the road.Breast implant extrusion can occur for various reasons, such as improper wound healing due to an infection, trauma, too little soft-tissue coverage, oversized implant coupled with too little tissue coverage, or lack of blood supply.Breast-implant extrusion calls for surgery and removal of the implant.Breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Most women undergoing augmentation or reconstruction with a mammary implant will experience post-operatory breast and/or chest pain.While this pain usually recedes in most women as they heal after surgery, it can become a chronic problem in other women.Hematoma, migration, infection, implants that are too large or capsular contracture can cause chronic pain.Sudden, severe pain may be associated with implant rupture.The surgeon should instruct the patient to immediately report if there is significant pain or if pain persists.A complete review of the dhr for lot involved was carried out, no deviation was found in the manufacturing process.Additionally, there were no indications of abnormalities in raw materials or manufacturing processes which may have affected this particular batch.Establishment labs will continue to monitor for similar complaints/trends.As part of the post market surveillance process, monitoring of the primary endpoints reported is performed to determine unfavorable trends.Per our current complaint data report, no unfavorable trends were detected on this product.No further actions are required.
|
|
Argentina.Allegedly, in march, a patient underwent implantation surgery.A few days after the procedure, one of the implants became exposed.Due to this complication, the treating physician decided to reintervene surgically to evaluate the situation.During the surgical procedure, both implants were removed for assessment.Upon inspection, it was confirmed that the implant in the right breast was ruptured, with serial number sn (b)(6).No patient demographics, such as weight, or ethnicity, were provided by the reporter.Efforts to gather additional information are ongoing, and the investigation remains in progress.
|