Model Number 9500-161
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that signal loss occurred.Performance data was reviewed.The allegation was confirmed due to the finding of signal loss over an hour within the investigation window.The probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
232 s dobson rd
mesa AZ 85202
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key25047563
Report Number3004753838-2026-142645
Device Sequence Number15897902
Product Code QBJ
UDI-Device Identifier00386270004123
UDI-Public(01)00386270004123(11)250901(17)270228(10)1725246001
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9500-161
Device Catalogue NumberSTE-AT-030
Device Lot Number1725246001
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/11/2026
Initial Report FDA Received Date04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TANDEM T: SLIM X2 INSULIN PUMP.
Patient Age52 YR
Patient SexFemale
-
-