| Catalog Number |
CDH25A |
| Medical Device Problem Codes |
Failure to Form Staple (2579); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 4/30/2026.D4: batch # unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during the unk surgery.After fired, noted that a few staples formed with irregular shapes and anastomotic site was oozing.Suture was used to oversew.There was no patient consequence reported, no additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/21/2026.D4: batch # 808d62.Investigation summary: the product was returned to ees for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the cdh25a device arrived with no apparent damage.The breakaway washer was present and uncut, and the device was fully loaded with staples, indicating that the device had not been fired.The device was tested for functionality with the returned washer and it fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event described could not be confirmed as no malformed staples were observed during the functional test and the device was returned without detectable damage.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 808d62, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|