(b)(4).Date sent: 4/30/2026.D4: batch # 905d08.Investigation summary: the product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the cdh31p device arrived with no apparent damage with anvil missing.The breakaway washer was not present and the device was fully loaded with staples, indicating that the device had not been fired.No battery was received, however, when the test battery was installed, the green checkmark illuminated indicated that the device was not reset.No functional test was performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the actuator was touching the stop switch actuator.This defect has been correlated to a manufacturing process.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.Device history review: a manufacturing record evaluation was performed for the finished device batch number 905d08, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|