| Model Number |
3320 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported via literature article that burr was stuck in lesion.Slow flow is a frequent complication of rotational atherectomy (ra) for severe coronary artery calcification (cac), yet effective preprocedural risk stratification tools are lacking.The cha2ds2 vasc hsf score incorporates multiple cardiovascular risk factors and may help predict adverse procedural outcomes.(b)(6) patients were analyzed retrospectively with severe cac undergoing ra.Patients were divided into a slow flow group (thrombolysis in myocardial infarction (timi) 0 to 2) and normal flow group (timi 3) based on post-ra angiographic assessment.Clinical, procedural, and angiographic data were compared.Independent predictors were identified using multivariate logistic regression, and receiver operating characteristic analysis assessed the predictive performance of variables.Ra was performed using the rotablator system (boston scientific (bsc), usa), which includes the rotawire guidewire, burr, and advancer mechanism.Initially, a standard 0.014-inch non-bsc coronary guidewire was advanced across the calcified lesion to reach the distal vessel segment.A non-bsc microcatheter was subsequently used to facilitate exchange for a 0.009 inch rotawire, either floppy or extra-support, allowing for precise and controlled burr delivery.Burr size was determined based on angiographic and intravascular imaging findings, generally starting with a 1.25 to 1.5 mm burr, with escalation as needed.The burr was advanced over the wire to a location just proximal to the target lesion using dynaglide mode, which facilitates smooth entry.Rotational speed was maintained between 140,000 and 190,000 rpm and modulated according to procedural demands.During activation, a gentle pecking motion was used for burr advancement, with each pass limited to less than equal to 30 s.Operators closely monitored for signs of excessive deceleration (e.G., a speed drop greater than 5,000 rpm), which could indicate high resistance or burr entrapment, particularly in lesions involving the right coronary ostium.Upon successful lesion crossing, the burr was withdrawn using either the dynaglide technique or balloon trapping.To minimize thermal injury and the occurrence of slow flow, a cocktail containing nicorandil (12 mg), isosorbide dinitrate (2.5 mg), and heparin (10,000 units in 500 ml saline) was continuously infused through the ra system.This infusion protocol was uniformly applied to all patients included in the study, in accordance with a standardized institutional guideline, ensuring consistency in pharmacologic support during ra procedures.Intravenous heparin was administered separately to maintain an activated clotting time of greater than 250 s throughout the procedure.Following ra, adequate contrast injection was used to reassess coronary perfusion, and timi flow grading was performed immediately.Additional lesion preparation was conducted using non-compliant, scoring, or cutting balloons to optimize outcomes of stent implantation or drug-coated balloon therapy.In selected cases, halfway ra targeting only the proximal segment of a long-calcified lesion was employed, with the distal portion treated using balloon angioplasty.This approach was typically reserved for lesions with complex or severe angulation.For patients at elevated procedural risk (e.G., those with reduced lvef, unprotected left main stenosis, or triple-vessel disease), intra-aortic balloon pump (iabp) support was inserted via femoral access prior to ra to enhance hemodynamic stability and reduce the incidence of periprocedural complications, including slow flow and myocardial injury.Slow flow occurred in 18.8% of patients.The slow flow group had significantly higher cha2ds2 vasc hsf scores, longer lesion lengths, smaller reference vessel diameters, and lower preprocedural systolic blood pressure (sbp).Multivariate analysis identified all four as independent predictors.The cha2ds2 vasc hsf score alone had moderate predictive value (area under the curve auc is 0.658), while a combined model incorporating the score with lesion length, reference diameter, and sbp achieved the highest accuracy auc is 0.893.Slow flow was also associated with a higher in-hospital rate of major adverse cardiovascular events.The cha2ds2 vasc hsf score is an independent predictor of slow flow during ra and, when combined with anatomical and hemodynamic variables, may enhance risk stratification.These findings support its use in preprocedural planning for patients with severe cac.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: device history review: a review was not completed of the device history records (dhr) for the device.The dhr review cannot be performed as no batch number was reported and a ship history cannot be performed since the upn was not reported.Device analysis findings: the device was not returned for analysis; therefore, a technical analysis could not be performed.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review: a risk review was performed, and it confirmed that the event was defined in the risk documentation and is recorded accordingly in the product record review (prr) table.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: within the literature, there were no indications of device complications during testing or prior to attempted use inside the patient.The instructions for use include testing of the device before usage within the body and state that use of the device is to be discontinued if evidence of a failure or damage is present.As there is no indication of device issues prior to use, it was considered likely that the reported events were attributable to factors encountered during the procedure.Based on a thorough review of the reported complaint, the most probable cause for this complaint was considered adverse event related to procedure.B3 date of event: corrected, earliest possible date of event as the patient underwent pci (b)(6) 2016 to (b)(6) 2023 e1 initial reporter phone:(b)(6) detailed product information was not provided to bsc, because the product is unknown, we are unable to provide the unique identifier (udi) # and other specific product information.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event: earliest possible date of event as the patient underwent pci (b)(6) 2016 to (b)(6) 2023 e1 initial reporter phone: (b)(6).Wu, x., wu, m., huang, h.Et al.Predictive value of the cha2ds2 vasc hsf score for slow flow during rotational atherectomy in patients with severe coronary artery calcification.Bmc cardiovasc disord 25, 469 (2025).Https://doi.Org/10.1186/s12872-025-04931-1.
|
| |
|
Event or Problem Description
|
|
It was reported via literature article that burr was stuck in lesion.Slow flow is a frequent complication of rotational atherectomy (ra) for severe coronary artery calcification (cac), yet effective preprocedural risk stratification tools are lacking.The cha2ds2 vasc hsf score incorporates multiple cardiovascular risk factors and may help predict adverse procedural outcomes.256 patients were analyzed retrospectively with severe cac undergoing ra.Patients were divided into a slow flow group (thrombolysis in myocardial infarction (timi) 0 to 2) and normal flow group (timi 3) based on post-ra angiographic assessment.Clinical, procedural, and angiographic data were compared.Independent predictors were identified using multivariate logistic regression, and receiver operating characteristic analysis assessed the predictive performance of variables.Ra was performed using the rotablator system (boston scientific (bsc), usa), which includes the rotawire guidewire, burr, and advancer mechanism.Initially, a standard 0.014-inch non-bsc coronary guidewire was advanced across the calcified lesion to reach the distal vessel segment.A non-bsc microcatheter was subsequently used to facilitate exchange for a 0.009 inch rotawire, either floppy or extra-support, allowing for precise and controlled burr delivery.Burr size was determined based on angiographic and intravascular imaging findings, generally starting with a 1.25 to 1.5 mm burr, with escalation as needed.The burr was advanced over the wire to a location just proximal to the target lesion using dynaglide mode, which facilitates smooth entry.Rotational speed was maintained between 140,000 and 190,000 rpm and modulated according to procedural demands.During activation, a gentle pecking motion was used for burr advancement, with each pass limited to less than equal to 30 s.Operators closely monitored for signs of excessive deceleration (e.G., a speed drop greater than 5,000 rpm), which could indicate high resistance or burr entrapment, particularly in lesions involving the right coronary ostium.Upon successful lesion crossing, the burr was withdrawn using either the dynaglide technique or balloon trapping.To minimize thermal injury and the occurrence of slow flow, a cocktail containing nicorandil (12 mg), isosorbide dinitrate (2.5 mg), and heparin (10,000 units in 500 ml saline) was continuously infused through the ra system.This infusion protocol was uniformly applied to all patients included in the study, in accordance with a standardized institutional guideline, ensuring consistency in pharmacologic support during ra procedures.Intravenous heparin was administered separately to maintain an activated clotting time of greater than 250 s throughout the procedure.Following ra, adequate contrast injection was used to reassess coronary perfusion, and timi flow grading was performed immediately.Additional lesion preparation was conducted using non-compliant, scoring, or cutting balloons to optimize outcomes of stent implantation or drug-coated balloon therapy.In selected cases, halfway ra targeting only the proximal segment of a long-calcified lesion was employed, with the distal portion treated using balloon angioplasty.This approach was typically reserved for lesions with complex or severe angulation.For patients at elevated procedural risk (e.G., those with reduced lvef, unprotected left main stenosis, or triple-vessel disease), intra-aortic balloon pump (iabp) support was inserted via femoral access prior to ra to enhance hemodynamic stability and reduce the incidence of periprocedural complications, including slow flow and myocardial injury.Slow flow occurred in 18.8% of patients.The slow flow group had significantly higher cha2ds2 vasc hsf scores, longer lesion lengths, smaller reference vessel diameters, and lower preprocedural systolic blood pressure (sbp).Multivariate analysis identified all four as independent predictors.The cha2ds2 vasc hsf score alone had moderate predictive value (area under the curve auc is 0.658), while a combined model incorporating the score with lesion length, reference diameter, and sbp achieved the highest accuracy auc is 0.893.Slow flow was also associated with a higher in-hospital rate of major adverse cardiovascular events.The cha2ds2 vasc hsf score is an independent predictor of slow flow during ra and, when combined with anatomical and hemodynamic variables, may enhance risk stratification.These findings support its use in preprocedural planning for patients with severe cac.
|
| |
|