• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE", 5 MM, 30°, NTSC, AUTOCLAVABLE; RIGID VIDEO LAPAROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH VIDEO TELESCOPE "ENDOEYE", 5 MM, 30°, NTSC, AUTOCLAVABLE; RIGID VIDEO LAPAROSCOPE Back to Search Results
Model Number WA50023B
Medical Device Problem Codes Poor Quality Image (1408); Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause of dirt under objective lens could not be established.However, the dust reflection and distorted image occurred due to faulty cable.The date of event is unknown.A supplement report will be submitted if provided.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the rigid video laparoscope exhibited debris under objective lens, dust reflection in image and distorted image.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEO TELESCOPE "ENDOEYE", 5 MM, 30°, NTSC, AUTOCLAVABLE
Common Device Name
RIGID VIDEO LAPAROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25050276
Report Number9610773-2026-03429
Device Sequence Number16066661
Product Code HET
UDI-Device Identifier04042761066501
UDI-Public4042761066501
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K955456
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWA50023B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/29/2026
Initial Report FDA Received Date04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-