This case was reviewed and investigated according to the manufacture¿s policy.Blocks a2-a6: no patient involvement.Blocks b6 & b7: no patient involvement.Block c: not applicable.Block d4: expiration date is not applicable.Blocks d6, d7 & d10: not applicable; no patient involvement.Block h6: the damage to the touchscreen monitor and cable occurred at the customer site.The device integrity can be affected by external factors such as device manipulation, its impact, and applied pressure associated with use and handling.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
|
It was reported that an intrasight system touchscreen monitor was cracked on the corner and the cable broke at the connection point.There was no patient present and no user injury reported.The touchscreen monitor and cable were returned for evaluation.Visual inspection of the monitor confirmed a crack with missing screen material, resulting in sharp edges.For the cable, damage was confirmed on the connector.This product problem is being reported in an abundance of caution because the touchscreen monitor has sharp edges that can result in a potential for harm.
|