| Model Number |
P1190A |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found loose stopper inside caster.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Check the brakes to see whether the bed moves when the brake bar or the lh/rh brake pedals are pressed.Repair as necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in feb 5, 2026.It is unknown if the facility performed any other preventative maintenance on this bed.The technician adjusted stopper inside caster until it would make contact with wheel to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
Baxter received a report that advanta2 rental bed had brakes not holding while brake engaged.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|