• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number D-EVOLUTFX-2329
Medical Device Problem Codes Appropriate Device Problem Term/Code Not Available (3191); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Rupture (2208); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during the attempted implant of a transcatheter aortic valve using a left trans carotid approach, resistance was felt while advancing the delivery catheter system (dcs) towards the aortic arch.Bleeding was identified, and a complete rupture of the left carotid artery as identified.Subsequently, the dcs was withdrawn from the patient, and vascular repair was performed.Subsequently, a new valve and dcs were opened and loaded, and the 14 french (fr) sentrant sheath was inserted into the patient.Prior to advancing the dcs, the operator was unable to withdraw the sheath.Rotational maneuvers and increased force were attempted, however the sheath remained stuck in the right iliac artery.Subsequently, the procedure was aborted due to risk of iliac rupture.It was later identified that the sheath had torn upon removal, and an external iliac artery rupture occurred.A balloon was inflated in the aorta to maintain hemodynamic control, and vascular repair was performed successfully.
 
Additional Manufacturer Narrative
Continuation of d10: product id {evfxplus-29}; product lot/serial number {(b)(6)}; product type: {0195-heartvalves}; implant date: {n/a}, explant date: {n/a} product id {sensh1428w}; product lot/serial number {(b)(6)}; product type: {0195-heartvalves}; implant date {non-implantable} select patient information cannot be included in the regulatory report due to regional privacy regulations.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during the attempted implant of a transcatheter aortic valve using a left trans carotid approach, resistance was felt while advancing the delivery catheter system (dcs) towards the aortic arch.Bleeding was identified, and a complete rupture of the left carotid artery as identified.Subsequently, the dcs was withdrawn from the patient, and vascular repair was performed.Subsequently, a new valve and dcs were opened and loaded, and the 14 french (fr) sentrant sheath was inserted into the patient.Prior to advancing the dcs, the operator was unable to withdraw the sheath.Rotational maneuvers and increased force were attempted, however the sheath remained stuck in the right iliac artery.Subsequently, the procedure was aborted due to risk of iliac rupture.It was later identified that the sheath had torn upon removal, and an external iliac artery rupture occurred.A balloon was inflated in the aorta to maintain hemodynamic control, and vascular repair was performed successfully.Additional information was received that per the physician, the cause of the carotid artery rupture was not specified; however, resistance was felt when crossing the aortic arch with the "dis", and calcium was noted at the ostia of the artery.Additionally, the dcs did not cause or contribute to the carotid artery rupture.The minimum diameter of the access vessel at the right femoral artery was 3.5 millimeters (mm), and the medtronic sheath was inserted into the right femoral artery.Per the physician, the cause of the iliac artery rupture was during extraction, the physician was unable to remove the medtronic sheath as it was stuck in the iliac artery; however, the dcs did not cause or contribute to the iliac artery rupture, and a "dis" was not used.
 
Additional Manufacturer Narrative
Updated data: b5.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received that ¿dis¿ meant delivery catheter system (dcs).It was further reported that the second dcs was never inserted into the femoral artery.
 
Additional Manufacturer Narrative
Additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Updated data: b5, h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX DCS
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key25056197
Report Number9612164-2026-02271
Device Sequence Number15791197
Product Code NPT
UDI-Device Identifier00763000942700
UDI-Public(01)00763000942700(17)270922(10)0013034291(20)01
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberD-EVOLUTFX-2329
Device Catalogue NumberD-EVOLUTFX-2329
Device Lot Number0013034291
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/29/2026
Supplement Date Received by Manufacturer05/03/2026
05/04/2026
Initial Report FDA Received Date05/01/2026
Supplement Report FDA Received Date05/04/2026
05/05/2026
Date Device Manufactured09/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-