| Model Number |
BF-XT190 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.B3: date of event is unknown.Additional information will be provided if available.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: stress problem led to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The bronchovideoscope was returned to the service center with a report of internal damage.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, suction flow was low due to obstruction was found.There was no patient involvement.
|
| |
|