Model Number ENF-GP
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
The rhino laryngofiberscope was returned to the olympus service center with a report of leaking at the light source.This does not indicate a medical device regulations reportable event.However, a reportable failure was found during testing at the olympus service center.During inspection of the device, looseness of the light guide cover glass was found due to a cause linked to the device but unable to trace more specifically.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause linked to the device but unable to trace more specifically.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RHINO-LARYNGOFIBERSCOPE
Common Device Name
RHINO-LARYNGOFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25057184
Report Number3002808148-2026-14118
Device Sequence Number18466673
Product Code EOB
UDI-Device Identifier04953170078231
UDI-Public04953170078231
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K011869
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberENF-GP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/21/2026
Initial Report FDA Received Date05/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-