SMITH & NEPHEW, INC. HEALICOIL KNOTLESS RGNST; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72205135 |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference number:(b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that during a fdl transfer, the threads of two (2) healicoil knotless anchors were not advancing into the bone, and the distal threads broke.All the components were removed from the patient, using tweezers, and a screw was used instead.Additional anesthesia was not required as consequence of the surgical prolongation.The procedure was resumed, without any delay, using a similar s+n back-up product in the originally drilled bone hole.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|