MEDLINE INDUSTRIES, LP MEDLINE; ISO-SHIELD CIRCUIT,INFNT,HTD DL,DRAIN Back to Search Results
Catalog Number HUD88015KIT
Medical Device Problem Codes Material Discolored (1170); Melted (1385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2026
Type of Reportable Event Malfunction
Event or Problem Description
According to the facility, on (b)(6) 2026, the circuits were being used on neonatal patients requiring respiratory support in critical care, including intubation and noninvasive positive pressure ventilation.The heater portion began alarming "low temp" and upon investigation, the circuits were found melted and turning black at the connection feeding to the heater.The heater was turned off, unplugged, and the circuit was removed and replaced.The facility reports no medical intervention or treatment required for the patients, they are in "stable condition." no additional information at this time.
 
Additional Manufacturer Narrative
According to the facility, on (b)(6) 2026, the circuits were being used on neonatal patients requiring respiratory support in critical care, including intubation and noninvasive positive pressure ventilation.The heater portion began alarming "low temp" and upon investigation, the circuits were found melted and turning black at the connection feeding to the heater.The heater was turned off, unplugged, and the circuit was removed and replaced.The facility reports no medical intervention or treatment required for the patients, they are in "stable condition." no additional information at this time.It has been determined that the reported event could cause or contribute to serious injury requiring medical intervention, therefore, this medwatch is being filed.If any further relevant information is identified or obtained, a supplemental medwatch will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ISO-SHIELD CIRCUIT,INFNT,HTD DL,DRAIN
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes dr
northfield IL 60093
Manufacturer Contact
gavin-ainsley nadong
three lakes dr
northfield, IL 60093
MDR Report Key25060148
Report Number1417592-2026-00839
Device Sequence Number18498528
Product Code OGG
UDI-Device Identifier10197344010479
UDI-Public10197344010479
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHUD88015KIT
Device Lot Number25DOA334
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2026
Initial Report FDA Received Date05/01/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/07/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Patient Weight2 KG
-
-