Model Number 797415
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an intrasight mobile system touchscreen monitor was cracked.There was no patient present and no user injury reported.The touchscreen monitor was returned for evaluation.Visual inspection confirmed a crack with missing screen material, resulting in sharp edges.This product problem is being reported in an abundance of caution because the touchscreen monitor has sharp edges that can result in a potential for harm.
 
Additional Manufacturer Narrative
This case was reviewed and investigated according to the manufacture¿s policy.Blocks a2-a6: no patient involvement.Blocks b6 & b7: no patient involvement.Block c: not applicable.Block d4: expiration date is not applicable.Blocks d6, d7 & d10: not applicable; no patient involvement.Block h6: based on the returned screen, the probable cause is likely damaged from use.The device integrity can be affected by external factors such as device manipulation, its impact, and applied pressure associated with use and handling.Blocks h7, h9 & h10: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRASIGHT MOBILE SYSTEM
Common Device Name
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
VOLCANO CORPORATION
3721 valley centre drive #500
san diego CA 92130
Manufacturer (Section G)
VOLCANO CORPORATION
3721 valley centre drive #500
san diego CA 92130
Manufacturer Contact
danielle bullock
3721 valley centre drive #500
san diego, CA 92130
858720-4125
MDR Report Key25060902
Report Number3008363989-2026-00072
Device Sequence Number18667104
Product Code IYO
UDI-Device Identifier00845225012915
UDI-Public(01)00845225012915(11)230228(10)1430965059
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K203719
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number797415
Device Catalogue Number797415
Device Lot Number1430965059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2026
Initial Report FDA Received Date05/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/28/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-