| Model Number |
LF-V |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The visera tracheal intubation videoscope was returned to the service center with a report of failed leak test.This does not indicate an mdr reportable event.However, mdr reportable failure was found during testing at the service center.During inspection of the device, the glue on the light guide lens was chipped was found.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for inspection.Based on the results of the investigation and the information provided it is likely due to some kind of stress problem.The most probable cause is component failure.The root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|