H10: internal complaint reference: (b)(4).H11:this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
|
|
It was reported that, during cori assisted tka surgery, when burring the distal femur, the purple removal icon was selected.It removed the purple model from the distal femur model on the cut screen.The surgeon requested the purple to be put back, but even when moving the resection +1mm on the planning screen, and moving the femur medial a few clicks, the purple did not return.The procedure was performed, after a non-significant delay, using the same device.No further complications were reported.
|