| Catalog Number |
SSF-240 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported "silicone rupture implanted 5 months ago".This record is for the left side.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported "silicone rupture implanted 5 months ago".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b3, d4.
|
| |
|
Event or Problem Description
|
|
Patient reported "silicone rupture implanted 5 months ago".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was impacted by emdr scheduled maintenance occurring (b)(6) 2026.Device evaluation:
per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ rupture: not observed through visual inspection.None of the other observations performed during the device analysis deformation and creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: d9, h3, h6.
|
| |
|
Search Alerts/Recalls
|