• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. MARQUEE; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. MARQUEE; BIOPSY INSTRUMENT Back to Search Results
Catalog Number MQ1816
Medical Device Problem Codes Difficult to Remove (1528); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.Two electronic photos were received and reviewed.In the first photo, one marquee needle is shown, along with truguide and coaxial needle and no bent was noted, and the second photo shows a technician holding a marquee needle and bent was noted to the needle.Based on the photo review, the reported material twisted /bent can be confirmed for one device.However, due to the absence of supporting evidence, the reported bent issue for remaining two devices and difficult to remove remains inconclusive.Therefore, the investigation is confirmed for the reported bent needle issue for one device as bend were noted to the needle.However, due to the absence of supporting evidence, the reported bent issue for remaining two devices remains inconclusive and the investigation for difficult to remove will remain inconclusive as no objective evidence was provided for review.A definitive root cause for the alleged material twisted / bent & difficult to remove could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent an ultrasound guided prostate biopsy procedure via normal density tissue using the marquee instrument.During the procedure, the device needle was allegedly bent and had difficulty in removing the product from patient.No coaxial needle was used.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARQUEE
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CAREFUSION D.R. 203 LTD.
zona franca las americas
km 22-e-1
santo domingo 11606
DR   11606
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key25064768
Report Number2020394-2026-01068
Device Sequence Number9400169
Product Code KNW
UDI-Device Identifier00801741097201
UDI-Public(01)00801741097201(11)250419(17)280228(10)0001614541
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMQ1816
Device Lot Number0001614541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2026
Initial Report FDA Received Date05/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-