| Catalog Number |
ES8816 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the implant was removed and replaced due to not filling "[inflation issue]".The reservoir was not replaced.
|
| |
|
Event or Problem Description
|
|
According to the available information, the implant was removed and replaced due to not filling "[inflation issue]".The reservoir was not replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump and cylinders 1 and 2 were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.Partial separations, surrounded by abrasion, were noted on the exhaust tubing of cylinder 2.These were not sites of leakage.No functional abnormalities were noted with either cylinder 1 or cylinder 2.The information received indicated that the device was not filling, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|