| Model Number |
UM-S20-17S |
| Medical Device Problem Codes |
Gas/Air Leak (2946); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, the most probable cause of the complaint is d02 - cause traced to component failure expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device inspection, the ultrasonic probe exhibited a cut mark on probe¿s cable unit, and internal oil was leaked out.There were no reports of patient involvement.
|
| |
|