Visual inspection: during the investigation, scratches and damage to the housing surface possibly caused by a tool were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves operate within the permissible tolerance in their respective relevant positions.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, deposits were found inside.Results: based on our investigation results, we can determine visible deposits in both valves.The deposits had no effect upon the specifications at the time of the examination.The shuntsystem operated within all specifications.At the time of the investigation, it is not clear to us how the aforementioned malfunctions occurred.Organic material in the csf can temporarily influence the function and are known side effects in hydrocephalus therapy.The examination of the return has been completed with the drafting of this report.A defect at the time of release can be excluded.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|