A set was returned to erbe usa and forwarded to the manufacturer for further inspection/testing as needed.Nevertheless, we received similar reports of sets leaking at the distal end of the irrigation line and down scopes.At this time, two (2) primary causes of the issue have been determined: 1.Off-label use in that erbe medical llc sets are being used with other manufacturers port connectors (note: per the notes on use for the set's, only erbe accessories (i.E., erbe port connectors), etc.Are to be used with erbe tubing/cap sets.2.A change had been made to the set's connector.Currently sets are now being produced with the previous connector to minimize/eliminate leaking.Other contributing causes of leaking at the end of scopes are also being investigated.If the evaluation findings are different than what has already been reported and/or any other prominent causes are determined and/or remedies are implemented to resolve leaking, a follow-up mdr will be filed.
|
|
It was reported that the hybrid co2 tubing/cap set for olympus® scopes & co2 was involved in an incident during an esophagogastroduodenoscopy (egd).The set was used with a co2 insufflator [model eco 2 plus, part number (p/n) 2n100-068, serial number (s/n) (b)(6)], irrigation pump (model eip 2, p/n 103250-000, s/n (b)(6), and a port connector (p/n 20325-211).Specifically, when insufflating the stomach with co2, water sprayed from the distal end of the scope as if the pedal of the irrigation pump was activated.Upon troubleshooting the issue during the case, the co2 flow from the insufflator was reduced from high to medium which resolved the spraying/leaking problem.There was no report of any patient injury.
|