SURGICAL SPECIALTIES MÉXICO, S. DE R.L. DE C.V. HENRY SCHEIN; 900-7466 Back to Search Results
Model Number 900-7466
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported on (b)(6) 2026 that during use a part of the suture needle broke off into the patient's skin.An x-ray was done to locate the needle piece and it was successfully removed.No additional injuries or medical intervention reported.
 
Additional Manufacturer Narrative
Sterile samples from the reported lot, including the actual devices, were not returned for visual review or analysis.Without the actual reported devices, photos of the devices, or detailed information on the method used to remove the device from the packaging, preoperative preparation of the device, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.Based on our review of the device history record (dhr) for the subject device, there were no recorded nonconformances, deviations, or other manufacturing anomalies that could account for the reported failure.All production and inspection processes were completed in accordance with specified requirements, and no irregularities were identified that would have impacted the device's functionality or quality.Review of labeling: contraindications due to the gradual loss of tensile strength which may occur over prolonged periods in vivo polyamide (nylon) suture should not be used where permanent retention of tensile strength is required.Warnings do not resterilize.Discard open, unused sutures and associated surgical needles.Users should be familiar with surgical procedures and techniques involving polyamide (nylon) suture before employing polyamide (nylon) suture for wound closure, as the risk of wound dehiscence may vary with the site of application and the suture material used.Physicians should consider the in vivo performance (under actions section) when selecting a suture for use in patients.The use of this suture may be inappropriate in elderly, malnourished, or debilitated patients, or in patients suffering from conditions which may delay wound healing.As with any foreign body, prolonged contact of any suture with salt solutions, such as those found in the urinary or biliary tracts, may result in calculus formation.Acceptable surgical practice should be followed with respect to drainage and closure of infected or contaminated wounds.Precautions care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.Infections, erythema, foreign body reactions, transient inflammatory reactions and in rare instances wound dehiscence are typical or foreseeable risks associated with any suture and hence are also potential complications associated with polyamide (nylon) suture.Acceptable surgical practice must be followed with respect to drainage and closure of infected or contaminated wounds.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point.Reshaping needles may cause them to lose strength and be less resistant to bending and breaking.Users should exercise caution when handling surgical needles to avoid inadvertent needle sticks.Discard used needles in ¿sharps¿ containers.As with any suture material, adequate knot security requires the accepted surgical technique of flat, square ties with additional throws as warranted by the surgical circumstance and the experience of the surgeon.The use of additional throws may be particularly appropriate when knotting monofilament sutures.Adverse reactions adverse effects associated with the use of this device may include: wound dehiscence; gradual loss of tensile strength over time; calculi formation in urinary and biliary tracts when prolonged contact with salt solutions such as urine and bile occurs; infected wounds; minimal acute inflammatory tissue reaction; and pain edema and erythema at the wound site.Broken needles may result in extended or additional surgeries or residual foreign bodies.Inadvertent needle sticks with contaminated surgical needles may result in the transmission of bloodborne pathogens.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HENRY SCHEIN
Common Device Name
900-7466
Manufacturer (Section D)
SURGICAL SPECIALTIES MÉXICO, S. DE R.L. DE C.V.
corredor tijuana rosarito 2000
#24702 b
tijuana baja california 22235
MX  22235
Manufacturer (Section G)
SURGICAL SPECIALTIES MÉXICO, S. DE R.L. DE C.V.
corredor tijuana-rosarito 2000
no. 24702-b
tijuana, b.c. 22235
MX   22235
Manufacturer Contact
virnalisa cobacha
2755 canyon boulevard
1st floor
boulder, CO 
MDR Report Key25071367
Report Number3010692967-2026-00024
Device Sequence Number10394035
Product Code GAR
UDI-Device Identifier00304040000106
UDI-Public00304040000106
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K903029
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number900-7466
Device Lot NumberF908ONF
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/04/2026
Supplement Report FDA Received Date07/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-