ZIMMER SURGICAL SA UNIVERSAL OSCILLATING SAW ATTACH; Saw, Powered, And Accessories Back to Search Results
Catalog Number 89-8509-450-60
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the oscillating saw has stopped working.No patient involvement.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4) g2: foreign - event occurred in poland.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the oscillating saw has stopped working.No patient involvement.No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; d4; d9; g3; h2; h4; h6 review of the most recent repair record identified the following related repair: malfunctioning excentric system, malfunctioning locking system, excentric shaft replaced and locking mechanism cleaned, device repaired and returned.Review of the previous repair record identified no related repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Review of complaint history found no additional related issues for this item and the reported part and serial combination.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL OSCILLATING SAW ATTACH
Common Device Name
Saw, Powered, And Accessories
Manufacturer (Section D)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
SZ  CH-1228
Manufacturer (Section G)
ZIMMER SURGICAL SA
3, ch. du pre fleuri
1228 plan-les-ouates
geneva CH-12 28
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key25072986
Report Number0008031000-2026-00032
Device Sequence Number10738603
Product Code HAB
UDI-Device Identifier00889024354913
UDI-Public(01)00889024354913(11)170403(10)5009651
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number89-8509-450-60
Device Lot Number5009651
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/30/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2026
Supplement Date Received by Manufacturer09/10/2026
Initial Report FDA Received Date05/04/2026
Supplement Report FDA Received Date09/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1