| Catalog Number |
89-8509-450-60 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the oscillating saw has stopped working.No patient involvement.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) g2: foreign - event occurred in poland.The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the oscillating saw has stopped working.No patient involvement.No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; d4; d9; g3; h2; h4; h6
review of the most recent repair record identified the following related repair: malfunctioning excentric system, malfunctioning locking system, excentric shaft replaced and locking mechanism cleaned, device repaired and returned.Review of the previous repair record identified no related repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Review of complaint history found no additional related issues for this item and the reported part and serial combination.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|