• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABIOMED, INC. - 1220648 IMPELLA; TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABIOMED, INC. - 1220648 IMPELLA; TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP Back to Search Results
Catalog Number 0052-3005
Medical Device Problem Codes Device Damaged Prior to Receipt by User (1570); Failure to Advance (2524); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device received damaged: the cause of device received damaged issue was not determined since no product/images were returned.Pd-any device-(twist, bent, kinked): the cause of pd-guidewire kink issue was not determined since no product/images were returned.
 
Additional Manufacturer Narrative
E4 was inadvertently omitted on the initial manufacturer device report.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
The complainant reported that their.018 guidewire kinked and was damaged coming out of package.They reported they had to open another.018 out of a different impella box.No patient harm has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPELLA
Common Device Name
TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP
Manufacturer (Section D)
ABIOMED, INC. - 1220648
22 cherry hill drive
danvers MA 01923
Manufacturer (Section G)
ABIOMED, INC.
22 cherry hill drive
danvers MA 01923
Manufacturer Contact
derek sammarco
22 cherry hill dr.
danvers, MA 01923
7814934967
MDR Report Key25073383
Report Number1220648-2026-07452
Device Sequence Number10695074
Product Code OZD
UDI-Device Identifier04260113630303
UDI-Public(01)04260113630303(11)250507(10)9464333
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0052-3005
Device Lot Number9464333
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2026
Supplement Date Received by Manufacturer05/13/2026
05/13/2026
06/22/2026
Initial Report FDA Received Date05/04/2026
Supplement Report FDA Received Date05/13/2026
05/27/2026
06/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age57 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-