Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The mega needle driver instrument was analyzed and found to have a frayed grip cable at the force amplification pulley.Some wires were broken, but the cable itself was not fully broken.There was no discoloration, corrosion, or contamination observed on the cable that would occur from improper flushing or rinsing during reprocessing.Common causes of frayed - distal instrument grip cables are attributed to damage to the cable through external collisions, during transport and/or storage, or during use or reprocessing.
|