| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The hearing performance of the device is affected.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
Since (b)(6) 2026, the user's hearing performance with the device has been affected.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.In addition, according to the implant registration card, a full insertion was not achieved at implantation surgery with one channel remaining extra-cochlear.However, to determine an exact root cause device investigation would be necessary.Re-implantation has been done but the explanted device has not been received for investigation yet.
|
| |
|