SYNTHES GMBH AO REAMING ATTACHMENT FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.210 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the reaming attachment device was frozen/will not move, could not engage the attachment ¿ coupling, the color band was chipping/fading, and the etching was illegible.It was further determined that the device failed pretest for general condition, markings, check the handpiece coupling, check pins length of handpiece coupling, and check general function in running mode.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in 2026.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned without an alleged deficiency.During routine service and repair, it was determined that the device was frozen/will not move.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
| |
|
Search Alerts/Recalls
|
|
|