| Model Number |
3401 |
| Medical Device Problem Codes |
Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
04/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) system delivered one shock as inappropriate therapy due to t-wave oversensing.Technical services (ts) was consulted and discussed stored data as well as device programmed settings.Ts discussed programming options as well as recommending capturing a reference subcutaneous electrocardiogram (s-ecg).This device remains in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A good faith effort was performed to obtain additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|