• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSCOR INC EMBLEM S-ICD; ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSCOR INC EMBLEM S-ICD; ELECTRODE Back to Search Results
Model Number 3401
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Read Input Signal (1581)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 04/09/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this subcutaneous implantable cardioverter defibrillator (s-icd) system delivered one shock as inappropriate therapy due to t-wave oversensing.Technical services (ts) was consulted and discussed stored data as well as device programmed settings.Ts discussed programming options as well as recommending capturing a reference subcutaneous electrocardiogram (s-ecg).This device remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
A good faith effort was performed to obtain additional information.At this time, no further information is available.Should additional information become available this report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Common Device Name
ELECTRODE
Manufacturer (Section D)
OSCOR INC
3816 desoto blvd.
palm harbor FL 34683
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key25082283
Report Number2124215-2026-24200
Device Sequence Number10395644
Product Code LWS
UDI-Device Identifier00802526575600
UDI-Public(01)00802526575600(17)180701
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/01/2018
Device Model Number3401
Device Catalogue Number3401
Device Lot NumberA135301
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2026
Supplement Date Received by Manufacturer06/25/2026
Initial Report FDA Received Date05/05/2026
Supplement Report FDA Received Date06/29/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
-
-