(b)(4).Date sent: 5/5/2026.D4: batch #unk.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.H4: the device manufacture date is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided; therefore, a device history could not be done.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
It was reported that during a robotic sleeve gastrectomy procedure, it was observed that the entire inner row (2d) on the remnant side of the stomach did not form.The 2nd fire blue with seamguard was fine, then it happened again with the 3rd fire blue (er60b) & seamguard.They decided to open a new stapler add the last two fires were blue (er60b) with seamguard and they were perfect.No delay or patient harm was reported.No additional information provided.
|