Catalog Number 0048-0024-JP
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6: updated codes based on the results of the investigation.H11: updated based on the results of the investigation.Investigation summary: failure to advance: the cause of the failure to advance was determined to be patient related due to the patient's tortuous anatomy.
 
Additional Manufacturer Narrative
This is one of two related reports.This report represent the impella cp and another report was submitted to represent the introducer.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
The complainant reported that a patient was to be implanted with an impella cp via percutaneous right femoral artery for mechanical circulatory support.Severe tortuosity made the impella insertion difficult.The insertion was abandoned.The patient's outcome is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPELLA
Common Device Name
TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP
Manufacturer (Section D)
ABIOMED, INC. - 1220648
22 cherry hill drive
danvers MA 01923
Manufacturer (Section G)
DANVERS
Manufacturer Contact
derek sammarco
22 cherry hill drive
danvers, MA 01923
7814934967
MDR Report Key25085612
Report Number1220648-2026-07522
Device Sequence Number18783042
Product Code OZD
UDI-Device Identifier00813502011371
UDI-Public(01)00813502011371(17)271111(10)2027821283
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P140003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0048-0024-JP
Device Lot Number2027821283
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2026
Supplement Date Received by Manufacturer06/17/2026
Initial Report FDA Received Date05/05/2026
Supplement Report FDA Received Date06/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-