• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CERENOVUS, INC. CEREGLIDE; CATHETER, THROMBUS RETRIEVER CATHETER, PERCUTANEOUS, NEUROVASCULATURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CERENOVUS, INC. CEREGLIDE; CATHETER, THROMBUS RETRIEVER CATHETER, PERCUTANEOUS, NEUROVASCULATURE Back to Search Results
Catalog Number NIC71137U
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.(b)(4).Information regarding patient identifier, date of birth, age, sex, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is nry/qjp.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (31684676) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no internal actions related to device manufacture or inspection.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during a stroke procedure, the physician used a 132cm cereglide 71 intermediate catheter (nic71132u / 31668075) through a zoom¿88 aspiration catheter (imperative care).Both catheters were well-hydrated (flushed), the cereglide reportedly got stuck in the zoom 88.During the advancement of the cereglide 71 catheter, the physician reported that it felt tight and as a result, the physician removed it and the cereglide 71 catheter stretched.The physician tried the second cereglide, a 137 cm cereglide 71 catheter (nic71137u / 31684676).Everything flushed well but the same issue was encountered.There was no negative impact on the patient.Additional information was received on 04-may-2026.Per the information, the procedure was targeting the m1 segment of the left middle cerebral artery.The target vessel was not excessively tortuous.The second cereglide catheter was also used with the zoom 88.The information indicated that for both cereglide catheters, the issue occurred on the first pass.After the second cereglide catheter, the physician used a zoom 7mm x 137mm to complete the procedure.The zoom 88 was also used to complete the procedure.It was confirmed that the zoom 88 was used again with the zoom 7mm x 137mm to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEREGLIDE
Common Device Name
CATHETER, THROMBUS RETRIEVER CATHETER, PERCUTANEOUS, NEUROVASCULATURE
Manufacturer (Section D)
CERENOVUS, INC.
6303 waterford district drive
suites 215 & 315
miami FL 33126
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
henry boden
31 technology dr
irvine, CA 92618
949788-2078
MDR Report Key25086331
Report Number3007628272-2026-00036
Device Sequence Number10537253
Product Code NRY
UDI-Device Identifier10886704085386
UDI-Public(01)10886704085386(17)280531(10)31684676
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberNIC71137U
Device Lot Number31684676
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2026
Supplement Date Received by Manufacturer05/21/2026
05/28/2026
Initial Report FDA Received Date05/05/2026
Supplement Report FDA Received Date05/21/2026
05/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-