| Model Number |
EV240152 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Failure to Convert Rhythm (1540)
|
| Health Effect - Clinical Code |
Tachycardia (2095)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient had recurrent ventricular tachycardia (vt) that could not be terminated with over twenty shocks.Additionally, oversensing noise was observed on the extravascular (ev) lead.Both the ev lead and the extravascular implantable cardioverter defibrillator (ev-icd) were explanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned and analyzed.Returned product analysis was performed and no anomalies were found.The analyst noted the full lead was returned with the anchoring sleeve.There were three suture marks on the anchoring sleeve.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|