D10 concomitant products: sigphandle, sig power sigphandle handle (serial # unknown); sigpshell, sig power sigpshell control shell (lot # unknown); sigadaptstnd, sig power sigadaptstnd linear adapter (serial # unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
According to the reporter, during the procedure, when setting up, while checking the opening and closing during connection, there was difficulty opening the jaw and the jaw would not open at all.The product was replaced with another new similar cartridge.Medtronic's initial evaluation of the incident found that there was stapling - incomplete retraction.The jaws of the reload were clamped.The i beam was slightly advanced.
|