| Model Number |
0174 |
| Medical Device Problem Codes |
High impedance (1291); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific and physical analysis could not be conducted.Investigation of the available information/data determined that the observed gradual rise in low-voltage shock impedance (lvsi) measurements was likely associated with calcification, as there was no conclusive evidence of compromised physical or electrical integrity of the lead.This condition may develop over time and can be reflected in a gradual and sustained increase in shock impedance measurements.Boston scientific issued a field safety notice to physicians in (b)(6) 2025 regarding the potential for reliance defibrillation leads with expanded polytetrafluoroethylene (eptfe) coating to exhibit a pattern of gradually increasing lvsi measurements due to calcification, which may result in elevated high-voltage shock impedance (hvsi) measurements and reduced shock efficacy.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
|
| |
|
Event or Problem Description
|
|
It was reported that during a clinic visit, the health care professional (hcp) called technical services (ts) for further review of the device data due to concerns of increase in right ventricular (rv) lead shock impedances.Upon review of the daily threshold and impedance measurements trends, the rv lead exhibited out of range intrinsic amplitude measurements and a gradual rise in the shock lead impedances from about 50 ohms to 108 ohms since 2018.It was noted the shock impedance measurements were within normal limits (wnl).Ts recommended for continued monitoring.At this time, this rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|