| Model Number |
ENF-V3 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed that during the evaluation, the rhino-laryngo videoscope exhibited that the a-rubber glue on the distal end glue is missing, peeled.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.In addition, the investigation found that the a-rubber glue on the distal end glue is missing, peeled.Based on the results of the investigation, the most probable cause of the discovered damage can be traced to component failure, which is expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor the field performance of this device.
|
| |
|