INTUITIVE SURGICAL, INC ENDOWRIST; POTTS SCISSORS Back to Search Results
Model Number 470001-12
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The potts scissors instrument was analyzed and found to have a broken distal pitch cable at the proximal clevis hole.The cable was fully broken.As a result of the full break, functional testing for motion related issues cannot be performed.There was no discoloration, corrosion, or contamination on the cable to suggest improper reprocessing as a contributing factor.Further inspection found no abnormally sharp or damaged components that could have contributed to the cable breaking.Common causes of broken - distal instrument pitch cables, whether partial or full, are attributed to damage to the cable through external collisions, during transport and/or storage, during use, or during reprocessing.
 
Event or Problem Description
It was reported that the customer experienced an issue with an intuitive product.The customer reported event has no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Common Device Name
POTTS SCISSORS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key25090151
Report Number2955842-2026-24010
Device Sequence Number13266414
Product Code NAY
UDI-Device Identifier00886874112120
UDI-Public(01)00886874112120(11)250220(10)K10250220
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470001-12
Device Catalogue Number470001
Device Lot NumberK10250220 0302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2026
Initial Report FDA Received Date05/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-