Model Number MR8-AS14
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bearings were damaged in the attachment prior to its use.At the time of the report, there was no patient impact.Additional information received indicated that detached bearings were found during evaluation.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation determined that the reported malfunction of bearing damage was confirmed.The likely cause of the f ailure was due to detached bearings.However, it was also noted that it was difficult to insert tool in the attachment, the bearings were found broken, corroded and the spring and washer was also found to be corroded.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX MR8
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key25092330
Report Number1625507-2026-00963
Device Sequence Number14722345
Product Code HBE
UDI-Device Identifier00643169912281
UDI-Public(01)00643169912281(11)191126
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AS14
Device Catalogue NumberMR8-AS14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2026
Initial Report FDA Received Date05/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1