• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. MAXCORE; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. MAXCORE; BIOPSY INSTRUMENT Back to Search Results
Catalog Number MC1820
Medical Device Problem Codes Self-Activation or Keying (1557); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigational summary: the physical device was not returned for evaluation.One electronic photo was provided and reviewed and photograph depicts the stylet alone, positioned in the half primed state.Therefore, the investigation is inconclusive with respect to the reported failure, as no objective evidence was provided for review.A definitive root cause for the reported failure to fire issue could not be determined based upon the provided information.It is unknown whether patient and/or procedural issues contributed to the reported event.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.Section a through f : the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent an ultrasound guided prostate biopsy procedure through normal density tissue using the maxcore instrument.During the procedure, the outer cannula could not be fired.No coaxial needle was used.The procedure was completed by another device.Further re was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXCORE
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key25092334
Report Number2020394-2026-01115
Device Sequence Number5079302
Product Code KNW
UDI-Device Identifier00801741084485
UDI-Public(01)00801741084485(11)250315(17)280131(10)REKP1230
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133948
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMC1820
Device Lot NumberREKP1230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2026
Initial Report FDA Received Date05/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1