This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.The device history record for lot 30247694 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.The stylet was received outside of the tube.The wire was bent and twisted.There is jagged tearing of the tube in the proximal portion, extending approximately 8cm from the base of the enfit connector.Some tears on the tubing are long.Due to the condition in which the stylet was returned, it was not possible to reinsert it into the tube.The stylet twisted wire does not exhibit breaks or fraying.The distal loop is formed and does not exhibit damage or deformation.A root cause has not been established.All information reasonably known as of 06 may 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Avanos medical inc.Received a single report that referenced two different incidents, which were associated with separate units, involving one patient.This is the first of two reports.Refer to for the first report.Refer to 9611594-2026-00282 for the second report.It was reported that a nasogastric tube was placed without difficulty.When the guidewire was being removed, resistance was felt and it was impossible to remove.After forcing, it did come out but it broke the tubing.
|