MAUDE Adverse Event Report: MDT PUERTO RICO OPERATIONS CO, MED REL CD HORIZON® SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT PUERTO RICO OPERATIONS CO, MED REL CD HORIZON® SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 559200029
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient who had revision surgery (t2 to l4 posterior spinal fusion) due to screw pull out.It was reported that during the procedure, one of the l4 screws was found to have pulled out of the bone.Additionally, four tulip heads were found to be loose on the rod despite the set screws being in their final tightened position.One of the four set screws had broken off and could not be loosened.The torque limit breakoff piece was broken off as it should be when following the final tightening surgical technique.The remaining three screws had shanks that were tight in the bone; however, their tulip heads could still be mobilized despite final tightened set screws.The loose tulips were discovered in interop during the revision.There were images showing one screw and one cable pulled out six days after initial surgery.The affected components were partially explanted.There were no patient symptoms reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON® SPINAL SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key25096189
Report Number2647346-2026-00236
Device Sequence Number14694568
Product Code NKB
UDI-Device Identifier00763000972776
UDI-Public(01)00763000972776(17)310310(10)H6132124
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K221244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number559200029
Device Catalogue Number559200029
Device Lot NumberH6132124
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2026
Supplement Date Received by Manufacturer06/15/2026
Initial Report FDA Received Date05/06/2026
Supplement Report FDA Received Date07/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-