A healthcare professional reported that this was a mechanical thrombectomy to treat an acute ischemic stroke (ais).During advancement of the emboguard 87, 95 cm (product code: bg8795u, lot number: 10293947) did not advance fully to the internal carotid artery (ica) and got kinked when the inner device was withdrawn.After the emboguard removal, replaced with a new emboguard and changed the method of advancing.The procedure was completed without further problems.There was no negative impact to the patient.It is unknown if a continuous flush was done.Puncture site was the groin.The kinking occurred during removal of the inner catheter.Location of catheter tip: common carotid artery (cca).The devices were used according to the instructions for use (ifu).Performed air removal according to ifu.Additional event information received on 06-may-2026 indicated that there was no allegation of patient injury due to an alleged product issue.
|
|
Manufacturer ref# (b)(4) information regarding patient weight, height, medical history, race, and ethnicity was not reported.Since no device has been received for analysis, no product investigation can be performed and the reported failure could not be evaluated.A device history review (dhr) associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.The instructions for use (ifu) of the emboguard devices cautions users against advancing or withdrawing the catheter or dilator against resistance without careful assessment of the cause of resistance using fluoroscopy.If cause cannot be determined, withdraw the catheter/dilator while exercising caution.Movement against resistance may result in damage to vessel or catheter/dilator, e.G.Kinking, or cause patient injury.Additionally, if the emboguard catheter becomes severely kinked, users are instructed to withdraw the entire system, (i.E.Emboguard catheter, intermediate catheter, guidewire, etc.).Use of a damaged device may result in vessel injury.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|