RESPIRONICS, INC. TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number IT1054096B
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported failure of cell undervoltage is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
 
Event or Problem Description
The trilogy 100 ventilator was returned to the third-party service center for remediation.The device was not in use on a patient at the time of the occurrence.During the evaluation of the device error codes in relation to cell undervoltage was observed in the device event log.The technician confirmed that there were no functional issues detected and no parts were replaced in regard to the error codes observed.The device was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25100682
Report Number2518422-2026-011912
Device Sequence Number4864965
Product Code CBK
UDI-Device Identifier00606959042933
UDI-Public(01)00606959042933
Combination Product (Y/N)N
Initial Reporter CountryRO
PMA/510(K) Number
K083526
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberIT1054096B
Device Catalogue NumberIT1054096B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2026
Initial Report FDA Received Date05/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/30/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-