• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV MARKSMAN; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV MARKSMAN; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number FA-55150-1030
Medical Device Problem Codes Retraction Problem (1536); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: product id ped-425-18 (b)(4); product id ped-400-20 (b778706); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received a report that the patient was undergoing treatment for amorphous, unruptured multiple aneurysms at the left c7 segment aneurysm with a max diameter of 5mm and a 4 mm neck diameter.It was noted that the patient's vessel tortuosity was moderate.The landing zone was 3.4 mm distally and 4.15 mm proximally.Dual antiplatelet treatment (dapt) was administered.The pru level was normal.It was reported that the middle segment of the ped-425-18 stent failed to open and remained unable to open despite repeated attempts.Another ped-400-20was used, but the distal tip end of the second stent failed to open, and the rest of the stent could not be retrieved.It had to be withdrawn.The angiographic result post procedure showed blood flow was unobstructed, with significant contrast stagnation within the aneurysm.The pipeline was used for an indication that is approved (on-label).The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.Ancillary devices include a u-track 5f sheath, marksman microcatheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARKSMAN
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc CA 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc CA 22210
MX   22210
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key25104534
Report Number2029214-2026-00828
Device Sequence Number13110492
Product Code KRA
UDI-Device Identifier00847536026261
UDI-Public(01)00847536026261(17)280604(10)231338907
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091559
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFA-55150-1030
Device Catalogue NumberFA-55150-1030
Device Lot Number231338907
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2026
Supplement Date Received by Manufacturer05/08/2026
Initial Report FDA Received Date05/06/2026
Supplement Report FDA Received Date06/05/2026
Date Device Manufactured06/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...
Patient Age48 YR
Patient SexFemale
Patient Weight55 KG
-
-